Input Power
|
115/230VAC 50/60Hz auto-select operation. Fused
at 3.3/1.8A.
|
Beam Energy
|
0 - 5000V, 1mA switch mode supply continuously
variable.
Output capacitance: 0.0047µF.
|
| Condenser Focus |
0 - 5000V, 1mA switch mode supply continuously
variable. Output voltage scales with energy. Output capacitance:
0.0047µF
|
Objective Focus
|
0 - 5000V, 1mA switch mode supply continuously
variable. Output voltage scales with energy. Output capacitance:
0.0047µF
|
Filament Power
|
Emission regulated supply with front panel
selectable filaments providing 5V@ 5A max.
|
Electron Bombardment
|
Electron accelerating voltage internally adjustable
to 150V.
Seven settings of electron emission current selectable from front
panel rotary switch.
|
Ion Extraction
|
Internally adjustable to 1500V.
|
Deflection
|
Variable bi-polar 350V DC supply for +X, -X,
+Y and -Y deflection. Remaining octupole elements are supplied from a resistive
divider network.
|
Interlocks
|
HV cable disconnection turns off HV supplies.
Adjustable high pressure interlock switches off HV supplies in the event
of system overpressure. System and Auxiliary interlocks provide total shutdown
in the event of system or auxiliary equipment failure
|
Front Panel Monitoring
|
Digital panel meters provide precision monitoring
of all critical parameters including; lens voltages (4 1/2 digits), ion
source pressure and beam current (3 1/2 digits), filament current and voltage
(3 1/2 digits), emission current (3 1/2 digits).
|
Chassis Dimensions
|
483(W)x132.5(H)x435.4(D) mm. 19 inch rack-mountable
desktop case 3U high.
|